4.4 Cleaning
Precleaning
All individual parts must be pre-cleaned.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
Manual cleaning
EQUIPMENT:
– Drinking water temperature of at least 40 °C
– Standard washing-up liquid
– Receptacle with at least 3 l capacity
PROCEDURE:
• Add about 1 teaspoon of wash-
ing-up liquid to 3 l warm drink-
ing water.
• Place all the parts in the wash-
ing-up water.
Application time: 5 minutes
• Occasionally move the parts back and forth.
• If there is visible soiling, use a medium-soft brush (such as a
toothbrush) which is used exclusively for this purpose.
RINSING:
• Rinse all individual parts thoroughly under running drinking
water at approx. 15 °C for 3 minutes.
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface, and
allow them to dry completely.
2) Validated with Palmolive
– 20 –
2
®
.