Changing The Battery - Cochlear Osia 2 Guide Rapide

Masquer les pouces Voir aussi pour Osia 2:
Table des Matières

Publicité

Changing the battery

The Osia 2 Sound Processor uses high power 675 (PR44) zinc air
disposable batteries designed for hearing implant use.
1
Hold the sound processor with the cover facing upwards.
2
Open the battery door until it is completely open.
3
Remove the old battery. Dispose of the battery according to
local regulations.
Remove the new battery from the pack. Remove the sticker on
the + side of the new battery and let it stand for a few seconds.
4
Insert the new battery with the + sign facing upwards in the
battery door.
5
Gently close the battery door until it is completely closed.
CAUTION
The sound processor must only be powered by a high power
675 (PR44) Zinc-Air battery for hearing implant use. If a
standard 675 battery is used, the device will not function.
Locking the battery door
You can lock the battery door to prevent it from opening
accidentally (tamper-proof). This is recommended when the sound
processor is being used by a child. To lock the battery door, close
the battery door completely and place the Tamperproof tool into
the battery door slot. Slide the locking pin up. To unlock the battery
door, place the Tamperproof tool into the battery door slot. Slide
the locking pin down into place.
Identification for bilateral users
You can use coloured stickers (red for right, blue for left) to make
identifying left and right processors easier. The identification
markers are placed on the bottom of the magnet prior to inserting
it into the device and will be visible through the transparent bottom
of the sound processor.
Connecting the programming cable
For programming the Osia 2 Sound Processor the Cochlear
Programming cable CS45 is used. To attach the programming
cable, the cover needs to be removed or replaced with the
programming cover. When attaching the programming cable to
the sound processor, align the dot on the cable with the dot on the
programming cover prior to inserting the cable.
NOTE
During programming mode, interference from nearby
electrical devices may disrupt communications between the
sound processor and the PC. If this occurs, the connection
status icon on the screen will inform the clinician of a loss of
communications. The clinician must then reconnect in the same
programming session in order to resume communication with
the sound processor.
6
1
4
R
L
Cochlear™ Osia
2 Sound Processor Clinician's Quick Guide
®
3
2
5
L
L
R

Publicité

Table des Matières
loading

Table des Matières