To remove a magnet
3
Place the magnet tool over the magnet.
4
Turn counter-clockwise to loosen the magnet.
5
Lift the magnet and pull out.
WARNING
Changing to a stronger magnet may cause a sensation of
pressure or skin irritation. To relieve this you can attach a
Cochlear SoftWear Pad to the back of the sound processor.
See the Osia 2 Sound Processor User Manual for details.
Changing the cover
To remove the cover
1
Open the battery door.
2
Press and lift to remove the cover.
To attach the cover
3
Place the cover over the upper part of the sound processor
base unit. The button should be visible through the cover
opening.
4
Press down on the cover around the button until a "click" is
heard.
5
Press down on the cover between microphone ports until a
"click" is heard.
6
Close the battery door.
NOTE
Double check no gaps are present between the cover and the
base.
Attaching a Cochlear SoftWear Pad
NOTE
The Cochlear SoftWear Pad is optional. If the patient
experiences discomfort when wearing the processor, they
can attach this adhesive pad to the back of the processor.
NOTE
A stronger magnet might be needed after attaching a
Cochlear SoftWear Pad.
1
Peel off the single backing strip on the adhesive side of the
pad.
2
Attach the pad to the back of the processor – press down
firmly.
3
Peel off the two semicircle backing covers on the cushion side
of the pad.