• Shift socket posterior (or foot anterior).
FOOT COVERS AND SPECTRA SOCKS (Figure 4)
• The device is designed to be used with a recommended foot cover
and Spectra sock.
• Use a shoehorn when fitting and removing foot covers to avoid
damage (Figure 4.1 and 4.2).
• Ensure the device is placed completely into the foot cover when
fitting.
MAINTAINANCE & CLEANING
• The device is weatherproof, but not resistant to corrosion. The device
should not come into contact with salt water or chlorinated water. In
cases where it does, the device must be rinsed with fresh water
immediately, and dried with a lint-free cloth.
• The device is not designed to be used in extreme conditions such as
diving or jumping into water.
• If device components get wet, parts should be dried with a lint-free
cloth.
Debris between blades
• Noise may occur if sand or debris are present between blades.
• The prosthetist should remove the Spectra sock and foot cover from
the device, and use compressed air to remove sand or debris.
Note: The device should always be used with a foot cover and Spectra
sock to minimize dust and dirt from entering between carbon blades.

ENVIRONMENTAL CONDITIONS

• Environmental Operating Temperature: -15°C (-5°F) to 50°C (122°F).
• Operating Humidity: 10% - 95% Relative Humidity.
• Shipping and Storage Temperature: -20°C (-4°F) to 70°C (158°F).
LIABILITY
The manufacturer recommends using the device only under the specified
conditions and for the intended purposes. The device must be
maintained according to the instructions for use. The manufacturer is
not liable for damage caused by component combinations that were not
authorized by the manufacturer.
COMPLIANCE
This device has been tested
according to ISO 10328 standard
to two million load cycles.
Depending on the user's activity
this corresponds to a duration of
use of two to three years. We
recommend carrying out yearly
safety checks.
ISO
- "P" - "m"kg *)
*)
Body mass limit not to be exceeded!
For speci c conditions and limitations of use see
manufacturer's written instructions on intended use!