SHOE HORN (Figure 10)
When removing and installing the cover use shoehorn to avoid damage
to foot or cover.

ENVIRONMENTAL CONDITIONS

• Operating temperature: -15 °C (+5 °F) to 50 °C (122 °F).
• Operating humidity: 10-95% relative humidity
• Shipping and storage temperature: -20 °C (-4 °F) to 70 °C (158 °F).
Note: The device should always be used with spectra socks and foot
covers to minimize dust and dirt from entering between carbon blades. It
is important to fix the spectra sock around pyramid location in order to
prevent noise issues coming from sand entering between the plates.
WATERPROOF IF USED WITH CARBON TUBE/DIRECTLY LAMINATED
ONTO THE SOCKET
A waterproof device allows use in a wet and/or humid environment and
temporary (<30 mins.) submersion of up to 1m in fresh water. Exposure
to fresh water under these circumstances shall have no harmful effect.
Dry thoroughly after contact with fresh water. Fresh water: Includes tap
water. Excludes salt and chlorinated water.
The components used with the foot need to be waterproof as well.
WEATHERPROOF IF USED WITH PYRAMID/T CONNECTOR
A weatherproof device allows use in a wet and/or humid environment
but does not allow submersion. Fresh water splashing against the
enclosure from any direction shall have no harmful effect. Dry thoroughly
after contact with fresh water. Fresh water: Includes tap water. Excludes
salt and chlorinated water.
LIABILITY
The manufacturer recommends using the device only under the specified
conditions and for the intended purposes. The device must be
maintained according to the instructions for use. The manufacturer is
not liable for damage caused by component combinations that were not
authorized by the manufacturer.
COMPLIANCE
This component has been tested according to ISO 10328 standard to two
million load cycles. Depending on the amputee's activity this
corresponds to a duration of use of two to three years. We recommend
carrying out regular yearly safety checks
ISO
*)
Body mass limit not to be exceeded!
For speci c conditions and limitations of use see
manufacturer's written instructions on intended use!
In these cases, the test loads have been adapted adequately on the basis
of the specified load level.
- "P" - "m"kg *)
In the standard mentioned,
test levels (P) are assigned
to a certain maximal body
masses (m in kg). In some
cases, which are marked
with, no test level is
assigned to the product
related maximal body mass.