3.3 Electrode installation

3.3.1 Install the reference electrode
Install the reference electrode as shown in the illustrated steps that follow.
At illustrated step 6, carefully turn the collar to break the seal. Then, move the collar up and down
and turn it clockwise and counter-clockwise.
At illustrated step 7, push the collar down and turn the collar less than 1/4 of a turn to lock the collar.
When the collar is locked, the collar does not turn. If the collar is not locked, the KCl electrolyte will
flow too fast from the reference electrode into the measurement cell.
At illustrated step 12, make sure to connect the cable with the blue connector to the reference
electrode.
Keep the storage bottle and caps for future use. Rinse the storage bottle with deionized water.
12 English