INTEGRA Metering AMFLO SONIC Smart Instructions De Montage Et Mode D'emploi page 23

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

WARNING
The device/system may be under pressure.
Risk of severe injury!
• Carry out work only on non-pressurised systems.
• When working on the device/system, watch out for leaking medium.
• Work may only be performed by authorised specialists in accordance
with the applicable regulations.
• Use appropriate protective equipment, particularly safety goggles.
3.1 Installation
CAUTION
The device has to be installed without tension (no torsional or flexional stress).
If necessary, supports have to be used for the cable and the device.
NOTICE
A seal is placed on the device in the factory to prevent manipulation and unau-
thorised maintainence work. Due to improper use and installation of the device,
the seal may be destroyed what results in ending of the guarantee.
NOTICE
Please note the information on the type plate!
Preferentially the flow transmitter is horizontally installed with consideration of the flow di-
rection (arrow on the housing) (vertical installation is possible). To prevent the formation of
air bubbles or dirt at the sensors, they have to be horizontally aligned for this installation
(connection box must not be horizontally aligned sideways +- 15°). Seals, screws as well
as tightening torque have to be used according to the pipe system. Additional recommen-
dations and installation conditions see following sub-section.