Before returning the Gage for service...
1. Install new or newly recharged batteries in the proper
alignment as shown within battery compartment.
2. Examine the probe tip for dirt or damage. The probe tip should
move up and down freely. The plate surrounding the
probe tip should be smooth and free from burrs and foreign
material.
3. Perform a Hard Reset (pg. 5).
4. Place the metal shim over the glass plate and attempt a
measurement.
5. If issue is not resolved, Update (pg. 10) your PosiTector
gage body and re-attempt measurements.
IMPORTANT:
If these steps do not resolve the issue and the Gage must be
returned for service, please follow the instructions provided at
www.defelsko.com/service.