Returning For Service - DeFelsko PosiTector SPG Manuel D'instructions

Masquer les pouces Voir aussi pour PosiTector SPG:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Before returning the Gage for service...
1. Install new or newly recharged batteries in the proper
alignment as shown within battery compartment.
2. Examine the probe tip for dirt or damage. The probe tip should
move up and down freely. The plate surrounding the
probe tip should be smooth and free from burrs and foreign
material.
3. Perform a Hard Reset (pg. 5).
4. Place the metal shim over the glass plate and attempt a
measurement.
5. If issue is not resolved, Update (pg. 10) your PosiTector
gage body and re-attempt measurements.
IMPORTANT:
If these steps do not resolve the issue and the Gage must be
returned for service, please follow the instructions provided at
www.defelsko.com/service.
Limited Warranty, Sole Remedy
DeFelsko's sole warranty, remedy, and liability are the express
limited warranty, remedy, and limited liability that are set forth on
its website: www.defelsko.com/terms
This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in whole or part,
by any means, without written permission from DeFelsko Corporation.
DeFelsko, PosiTector, and PosiSoft are trademarks of DeFelsko Corporation registered in the U.S. and in other
countries. Other brand or product names are trademarks or registered trademarks of their respective holders.
Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not
responsible for printing or clerical errors.

Returning for Service

and Limited Liability
www.defelsko.com
© 2022 DeFelsko Corporation USA All Rights Reserved

Publicité

Table des Matières
loading

Table des Matières