Service And Maintenance; Metrological Control; Technical Data - GERATHERM GT-3020 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EN

servIce and MaIntenance

display Icons and troubleshooting
display
description
Please check the battery first before
No display
sending it for repair.
Battery is empty.
Measured value is above 43.9 °C and
therefore outside the measuring range.
Measured value is under 32.0 °C and
therefore outside the measuring range.
Electronic failure.
If this occurs repeatedly, contact the
manufacturer.

MetrologIcal control

It is recommended to have a metrological check on the thermometer
carried out by an authorized laboratory at least every 2 years. This
ensures the correct function and measurement accuracy.
dIsposal
When disposing of the thermometer, batteries and packaging, the
applicable country-specific regulations must be observed.
• Dispose of the packaging of your thermometer according to local
regulations.
• Remove the battery before disposing of the thermometer.
• Do not throw used batteries in the household waste, but dispose
of them as hazardous waste or at a battery collection point in
specialist shops.
• Dispose of the thermometer in accordance with local regulations
for electronic waste.
• Do not dispose of the battery or thermometer in a fire.
guIdelInes and ManuFacturer's declaratIon
regardIng electroMagnetIc coMpatIBIlIty
(eMc)
WarnIng:
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
EM disturbances can affect the measurement accuracy of the
thermometer.
WarnIng:
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the GT-3020, including cables
specified by the manufacturer. Otherwise, degradation of the perfor-
mance of this equipment could result.
The complete EMC appendices are available on request.

tecHnIcal data

Type of thermometer
Mode of operation
Transient time
Measuring range
Measuring accuracy
Display resolution
Unit
Power supply
Weight / dimensions
Degree of protection
Working conditions
Storage and transport
conditions
Life time
Additional functions
10
Medical thermometer, direct mode
intermittent
60 seconds
32.0 °C - 43.9 °C
± 0.1 °C (in the measuring range
34.0 °C – 43.0 °C)
± 0.2 °C (in the measuring range
32.0 to 33.9 °C and 43.1 to 43.9 °C)
0.1 °C
Degree Celsius (°C)
1 button cell 1.5 V LR41 or compara-
ble (4000 measurements)
Appr. 17g
133.5 mm x 22 mm x 11.5 mm
IP22
Temperature: +5 °C to +40 °C
Humidity: 15 % to 90 % rel. Humidity
Air pressure: 700 – 1060 hPa
Temperature: -25 °C to +60 °C
Humidity: 15 % to 90 % rel. Humidity
Air pressure: 700 – 1060 hPa
5 years
Self-test, memory, Beeper