8 Care and maintenance

8.1 Cleaning and drying

1.
Blow dust off the glass.
2.
Do not touch the glass with the fingers.
3.
Use only a clean, soft cloth for cleaning. If necessary,
moisten the cloth slightly with pure alcohol or a little
water.
en
NOTE Do not use any other liquids as these may
damage the plastic components.
8.2 Storage
Remove the appliance from its case if it has become wet.
The tool, its carrying case and accessories should be
cleaned and dried (at maximum 63°C / 145°F). Repack
the equipment only once it has dried completely and then
store it in a dry place.
Check the accuracy of the equipment before it is used
after a long period of storage or transportation.
Remove the batteries from the appliance before storing
it for a long period. Leaking batteries may damage the
appliance.
Observe the temperature limits when storing your equip-
ment, especially in winter / summer if the equipment is
kept inside a motor vehicle (-25°C to +60°C).
9 Troubleshooting
Fault
The tool cannot be switched on.
The tool can be switched on but
no laser beam is visible.
Automatic leveling doesn't func-
tion.
NOTE
If the steps listed here do not remedy the error, or if further errors occur, have the tool repaired at a Hilti Measuring
Systems Service Center.
16
All manuals and user guides at all-guides.com
8.3 Transport
Use the Hilti toolbox or packaging of equivalent quality
for transporting or shipping your equipment.
8.4 Hilti Measuring Systems Service
Hilti Measuring Systems Service checks the tool and, if
deviations from the specified accuracy are found, recal-
ibrates the tool and checks it again to ensure conformity
with specifications. The service certificate provides writ-
ten confirmation of conformity with specifications at the
time of the test.
The following is recommended:
1.
2.
3.
4.
Possible cause
The battery is exhausted.
The battery is inserted the wrong way
round (incorrect polarity).
The battery compartment is not
closed.
The temperature is too high or too
low.
The tool is set up on an excessively
inclined surface.
The tool should be checked at suitable intervals,
depending on the frequency of normal use.
The tool should be checked at least once a year by
a Hilti Measuring Systems Service Center.
The tool should be checked by a Hilti Measuring
Systems Service Center if it has been abused in any
way.
The tool should be checked by a Hilti Measuring
Systems Service Center before being used for par-
ticularly important work.
Having the tool checked by a Hilti Measuring Sys-
tems Service Center does not relieve the user of
his/her obligation to check the tool before and dur-
ing use.
Remedy
Replace the battery.
Insert the battery correctly.
Close the battery compartment.
Allow the tool to cool down or warm
up.
Set up the tool on the level.