Problem
The display "Er:X:16"
appears.
The Enter key (send/
print) is pressed and the
display "Er:X:71"
appears.
The Enter key (send/
print) is pressed and the
display "Er:X:72"
appears.
11.SERVICING
Cause / Solution
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 84).
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 84).
On leaving the factory, your seca scale has an accu-
racy of ±0.15% or better. To preserve this level of
accuracy, the product must be set up with care and
serviced regularly. Depending on how frequently the
scale is used, we recommended servicing at intervals
of 3 to 5 years.
ATTENTION!
Incorrect measurements as a result of poor
servicing
– Please only have servicing and repairs
performed by an authorized service partner.
– You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.