9. FUNCTION CHECK

10.TROUBLESHOOTING
Problem
No weight displayed
with a load on the scale.
0.00 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display "StOP"
appears.
The display "tEMP"
appears.
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview".
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 89.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 89, you may not use the device.
– Have the device repaired by seca service or
an authorized specialist dealer.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
The corresponding point has a fault.
- Inform the service department.
Maximum load has been exceeded.
- Remove the load from the scale.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.