. Maintenance
fuse replacement
Check before each use
Visually inspect the device for signs of me
chanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.
Technical inspections
Due to their design, the bearings and
joints of the ARTROMOT® K do not
require maintenance and all materials are
protected from corrosion, BUT: For safety,
the devices require regular maintenance.
To maintain the functional and operational
safety, check all components for damage
and loose connections at least once a year.
These checks should be performed by
persons with adequate training and experi
ence. Damaged and worn parts must im
mediately be replaced with original spare
parts by authorized staff.
Warning!
Patient hazard
equipment malfunction and damage
− Refer repair and maintenance to
authorized persons. Persons are
authorized after training by a special
ist trained and commissioned by the
manufacturer.
ORMED GmbH will make all docu
ments required for servicing, such as
circuit diagrams, parts lists, descrip
tions or calibration instructions, avail
able to authorized experts.
The inspections can be carried out by DJO
Technical Service within the framework
of a service agreement. Contact DJO for
details.
Other than that, the manufacturer
does not require any other regular
maintenance.
Note!
Regarding further technical or other
inspections or inspection intervals,
observe applicable local requirements,
such as IEC
, DGUV or similar
regulations to be observed by users of
medical equipment or electric devices.
Fuse replacement
Warning!
Patient hazard
equipment malfunction and damage
The replacement of fuses must be
referred to specialists as defined in DIN
VDE
or IEC
or other applicable
standards e.g. biomedical technicians,
electricians, electronics installers .
Fuses used must be
S <
,
: T A L
S >
,
: T A H
Vac
Vac.