. Setting the treatment values
Warning!
Patient hazard
Before therapy, a test run consisting of
several exercise cycles must be completed
without the patient. Then, repeat the
test run with the patient and check that
the movements do not cause any pain.
Note!
See also . and . as well as VIII!
. General information
on programming the
ARTROMOT® K
. You activate the programming mode
by briefly pressing the MENU button on
the control pendant.
. The various treatment parameters and
functions are allocated to three Standard
model or five Comfort model program
ming levels
per level .
To be able to program a parameter you
will have to access the corresponding
programming level. This is also done
with the MENU button. With each press
of the button you advance one level.
The code M , M , etc. that appears in
the middle of the display indicates the
programming level.
. You activate the treatment parameters
and functions with the
buttons below the display. The symbols
parameter buttons
above the
indicate the assigned parameters and
functions.
This is what happens when you press
one of the parameter buttons to select a
parameter:
− The corresponding symbol appears on
the display in a larger format.
− The set value is displayed.
− The symbol above the parameter but
ton is highlighted.
. With the +/− buttons plus/minus you
. Having programmed all parameters, press
. Then press the START button to start
parameter
change the displayed value. When you
press and hold the key, the value will
change at a higher rate.
Some of the special functions can only
be enabled and disabled. This is done by
pressing the corresponding parameter
button or with the +/− buttons. Active
parameters are identified with a check
mark in the circle next to the symbol.
the STOP button to save the values.
therapy.
Note!
− Refer to sections . and . for a
description of the parameters.
− To view the set parameter values,
press the corresponding parameter
button. However, this is only possible
when you press the STOP button first.
− To prevent accidental changes of the
parameter settings, you can lock the
buttons. To do so, simultaneously
press the + and
buttons for approx.
seconds.
Press both buttons again for approx.
seconds to unlock.
− Selecting the "New Patient" function
will automatically delete the data on
the patient chip card. When you have
finished programming the unit and
press the STOP button, the settings
will automatically also be saved to the
patient chip card.
− Emergency stop function: The
ARTROMOT® K will stop immediately,
when any of the buttons is pressed
during therapy. Patient treatment can
be resumed by pressing the START
button. The unit will automatically
change the direction of movement.
− The manufacturer recommends an
application period of up to 1 hour.