. IEC
For ARTROMOT® K or ARTROMOT® K Classic devices with serial numbers >
The product designation ARTROMOT® K used in the text below includes the
ARTROMOT® K and ARTROMOT® K Classic products with all their variants.
The ARTROMOT® K device is subject to
particular precautions regarding electro
magnetic compatibility EMC . The device
must be installed and put into service strictly
in compliance with the EMC directives set
forth in the accompanying documents.
Portable and mobile RF communication
systems may a ect the ARTROMOT® K
The ARTROMOT® K should not be used
adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary,
ARTROMOT® K should be observed to
verify normal operation in the configuration
in which it will be used.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.
WARNING!
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
. Electromagnetic emissions
Guidance and Manufacturer's Declaration
ARTROMOT® K is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT® K unit is
used in such an environment.
Emissions test
RF emissions to CISPR
RF emissions to CISPR
Harmonic emissions to
IEC
Voltage fluctuations/flicker
emissions to IEC
and IEC
If it is necessary to replace assemblies or
cables, only the manufacturer's original
parts must be used to ensure continued
compliance with EMC requirements
after repair. This requirement applies to
the power supply unit, cables and cable
lengths, drive unit consisting of the motor
and the control system, the programming
unit incl. the coiled cable and the connector.
With regards to the electromagnetic
compatibility, there is no maintenance
operation required during the expected
product life time!
Compliance
Group
Class B
complies with class
A requirements
complies
Electromagnetic Emissions
Electromagnetic environment
guidance
ARTROMOT® K uses RF energy
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
ARTROMOT® K is suitable for use
in all establishments, including do
mestic and those directly connected
to the public low voltage power
supply network that supplies buil
dings used for domestic purposes.
.
.