Télécharger Imprimer la page

3B SCIENTIFIC PHYSICS U189021 Instructions D'utilisation page 9

Equipement complémentaire rmn

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 13
Hold the yoke in place with your thumbs and
use your fingers to pull the magnet unit towards
the front till the yoke is between the two mag-
nets. Then take the entire unit out of the basic
unit (see Fig. 5).
Take the discs out of the sample chamber.
Fig. 5 Removing the magnet unit from the basic unit
6.2 Connection to control console
Insert the probe into the sample chamber in
such a way that it rests on the housing (see Fig.
6).
Connect the cable for the probe to the "Probe
In" socket of the control console, taking note of
the socket's slot.
Connect the coils to the "Coil" sockets on the
back of the control console.
Connect the control console to its power supply
via the "12 VAC/1A" socket.
Fig. 6 Basic unit with probe
6.3 Calibration and settings
6.3.1 Using an oscilloscope
Connect the "SIGNAL OUT" socket of the control
console to channel 1 of the oscilloscope and the
"FIELD OUT" output to channel 2 (see Fig. 8).
Set the oscilloscope as follows:
Channel 1: 0.5 V DC
Channel 2: 0.5 V DC
Time base: 5 ms
Trigger from a falling edge on channel 1.
6.3.2 Using 3B NETlog™
Connect the "SIGNAL OUT" socket of the control
console to input
and "FIELD OUT" to the input
Connect the 3B NETlog™ unit to a computer and
run the 3BNETlab
Create a new data record from the "Measure-
ment lab" menu and define the following pa-
rameters:
Input A: Field, Input mode VDC, Input range 2 V
Input B: Signal, Input mode VDC, Input range 2
V
Measurement interval: 500 μs (2 kHz)
Set to trigger from Input A with a falling edge
and a positive trigger point at about 10 to 20%.
Activate the "Oscilloscope" button and start
measuring.
The oscilloscope window will open.
6.4 Experiment procedure
Insert the glycerine sample (yellow top) into the
sample chamber (see Fig. 7).
Fig. 7 Basic unit with glycerine sample inserted
Set a frequency of about 13 MHz on the control
console (since the frequency knob is a 10-turn
potentiometer, it may need to be turned by mul-
tiple revolutions).
Set the sensitivity to medium and adjust if nec-
essary.
At the optimum setting, the LED can be seen to
flicker slightly. If the LED lights up fully, the signal is
overloaded.
3
IN
U
on the 3B NETlog™ unit
B
IN
U .
A
TM
software

Publicité

loading