Drive unit

Mount the drive on the support rod (1) with the clamping ring (4). The black screw collar
(9) must face inwards towards the heating bath and the electric power cable (6) must point
down.
Secure the clamping ring (4) with the drive unit by tightening the star grip (5) on the
support rod (1).
Insert the power cable with plug (6) in the correct socket on the rear of the heating bath
housing (2).
Screw the support rod (7) for the condenser in the threaded hole of the coupling (8).
2.

Vapor support tube and evaporation flask

Remove the black screw collar (9), the spring tension ring (10), the PTFE seal (11) and the
silicon ring (12) from the drive.
From the side with the red locking knob (14), insert the vapor tube (A) with the black
support ring (B) in the drive unit as far as the stop. Make sure the side outlet of the vapor
tube is parallel to the support rod (7).
From the bath side, slide the silicon ring (12) - curved
face first - on the vapor tube (A). Then slide the
PTFE seal (11) - sealing lip first - on the vapor
support tube as far as it will go.
Place first the screw collar (9) and then the spring
tension ring (10) on the neck of the evaporation flask
(C, D, E) or transition piece (F).
Slide the flange over the vapor support tube (A) and
into the drive up against the seal. Push the flask as
far as it will go and half tighten the screw collar.
To give the screw collar and flask sufficient hold,
press the fastening knob (13) to lock the gear.
3.

Condenser

end. Using the butt flange (G), screw the condenser unit (H) to the vapor tube (A). The red
locking button (14) must be tight in order to provide a secure hold for the condenser and
support rod.
The glass set is easier to mount when the condenser is in
upright position.
14
The support rod (7) should be vertical. If not, loosen the red
locking knob (14) and pull out the fastening knob (13) so that
the coupling (8) can be swivelled 180°. The fastening knob (13)
will then latch home again and the support rod will be in upright
position. Re-tighten the red locking knob (14).
Screw the butt flange (G) to the threaded socket on the side of
the vapor support tube (A) and insert the sealing ring (G1).
Place the clamp (15) on the support rod (7) and fasten the
condenser (H) to the support rod with the clamp. Make sure that
the condenser's water and vacuum connections are at the top
22