Standard Low Air Loss Mattress - Sunrise Medical Jay Dermafloat LAL 1000 Serie Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Covering the entire mattress assembly is a low friction (i.e. low shear
producing) 70-denier nylon comforter. The bottom surface of the comforter
is coated with a vapor permeable polyurethane film. This coating creates a
monolithic, osmotic membrane impermeable to liquids, air, and bacteria
though still permeable to water vapor.
Quilted onto the base of the comforter is a spun bonded polyester fiberfill,
which exhibits less frictional resistance to nylon than that of the patient's
skin. The significance of this is that when the patient moves, the comforter
tends to move relative to the air cushions rather than relative to the patient.
Wounds, which can be exacerbated by shearing are in this manner
significantly minimized, if not prevented.
quilting is that it acts to provide a diffusion layer by with the water vapor can
be drawn away from the patient. Concentrations of moisture will diffuse
throughout, the comforter thus resulting in more efficient evaporation and
subsequent reduced skin maceration.
Jay Dermafloat LAL 1000 Series
MASTER CONTROL UNIT FEATURES
·
Compact, lightweight, and easy to use.
·
Sleek and durable ABS plastic housing with bedside bracket for easy
placement on the footboard of bed.
·
Push button connector for quick disconnection and rapid deflation as
needed for CPR.
·
Dialed Soft/Firm Comfort Control.
STANDARD LOW AIR LOSS MATTRESS FEATURES
·
Horizontal, tubular cell construction.
·
Individual air cushions designed for maximum support area and pressure
distribution.
·
Mattress Replacement System, eliminating the compromising effects of
an existing mattress.
·
Dimensions 80" x 35" x 10" (2 m x 89 cm, x 25 cm) when fully inflated
(20 - 8" (20 cm) air cells plus a 2" (5 cm) foam base).
Jay Dermafloat LAL 1100
A secondary feature of this
3

Publicité

Table des Matières
loading