Table des Matières

Publicité

Use

This paragraph details the fundamental operating procedures.
Make all electrical and pneumatic connections before the use of the
system.
While heating the vacuum chamber, the temperature of the inlet
flange must not exceed 120 °C.
WARNING!
Never use the turbopump when the inlet flange is not connected to the
vacuum chamber. Do not touch the turbopump or any of its accessories
during the heating process. The high temperatures may cause burns.
Avoid impacts, oscillations or harsh movements of the pump when in operation.
CAUTION!
The bearings may become dam-aged. Use air or inert gas free from dust or
particles for venting the pump. The pressure at the vent port must be less than 2
bar (above atmospheric pressure). For pumping aggressive gases, these pumps
are fitted with a special port to allow a steady flow of inert gas (like N
pump bearing protection (see the appendix "Technical Information").
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal. Do not
use the pump in presence of explosive gases.
TV 301SF Navigator User Manual /
87-900-957-01 (H)
Instructions for Use

Use

, Ar) for
2
231/300
16

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-8917969-8822969-8823969-8972969-8973

Table des Matières