Installation

Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Figure 2
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The TV 301SF Navigator must be connected to a primary pump (see
"Technical Information").
TV 301SF Navigator User Manual /
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C (see the diagram pressure-
temperature in the appendix "Technical Information")
relative humidity: 0 – 95 % (non-condensing)
87-900-957-01 (H)
Instructions for Use

Installation