Télécharger Imprimer la page

Calibrate - ProMinent DULCOTEST BCR 1-mA-0.5 ppm Guide De Montage Et De Service

Publicité

Les langues disponibles

Les langues disponibles

Put the sensor into operation
6.1

Calibrate

CAUTION!
Following the replacement of a
membrane cap or electrolyte, you
must carry out a gradient calibra‐
tion
The slope has to be re-calibrated
at regular intervals to ensure per‐
fect operation of the sensor. The
calibration intervals are 1 - 4
weeks, depending on the quality
of the water
Avoid air bubbles in the sample
water. These could result in incor‐
rect metering. Air bubbles that
adhere to the membrane of the
sensor may result in too low a
reading and thus in over-
metering.
Please note the applicable
national guidelines for calibration
intervals
Prerequisites
Constant flow rate at the in-line
probe housing
Constant temperature of the
sample water
Equal temperature of sample
water and sensor (wait approx.
15 minutes)
The sensor has been run-in
Constant pH value
50
Zero point calibration
If the sensor is operated with a ProMinent
control device, then zero point calibration
is not generally necessary. Perform a zero
point calibration if you are working in the
measuring range 0.01 ... 0.5 mg/l or
working in other measuring ranges at the
lower end of the measuring range (< 10
%).
1.
Dip the sensor in a bucket of clean
water without disinfectant or oxidant
(e.g. common still mineral water)
2.
Stir with the sensor until the meas‐
ured value at the control device is
stable for 5 mins
3.
Calibrate the control device to zero
in accordance with its operating
manual
4.
Re-install the sensor into the in-line
probe housing (e.g. DGMA; DLG III)
Gradient calibration
1.
Immediately determine the disinfec‐
tant content of the sample water
using a suitable measuring kit (e.g.
DPD)
2.
Set the value ascertained on the
control device according to its oper‐
ating instructions
ð Repeat the calibration after one
day to ensure that the sensor
has reached its maximum sen‐
sitivity (gradient).

Publicité

loading

Ce manuel est également adapté pour:

Dulcotest bcr 1-ma-2 ppmDulcotest bcr1-ma-10 ppm