Test a reference sample
1
Place electrode into the reference sample.
2
Press Start to begin the reference measurement and wait until the measurement display
freezes.
— or —
3
Press READ to manually end the measurement.
4
Press Save to enter the measured mV value as offset into the meter.
6.5  E ndpoint formats
Auto
With the automatic endpoint the selected stability criterion determines the end of an individual
read i ng depending on the behavior of the sensor used. This ensures an easy, quick and precise
mea s ure m ent.
1
Place sensor in the sample.
2
Press READ.
A appears on the display.
The measurement ends automatically when the measured value is stable. appears.
If READ is pressed before the signal is stable, the endpoint format changes to manual .
Manual
Unlike Auto, user interaction is required to stop the measurement reading in manual mode.
1
Place sensor in the sample.
2
Press READ.
M appears on the display.
appears on the display to signalize measurement stability.
3
Press READ to end the measurement. appears.
Timed
The measurement stops after the set time, which can be set between 5 s and 3600 s.
1
Place sensor in the sample.
2
Press READ.
T appears on the display.
appears on the display to signalize measurement stability.
The measurement ends automatically when the set time period expires. appears.
If READ is pressed before the signal is stable, the endpoint format changes to manual .
Information on the display
The following symbols appear in the display, depending on the endpoint setting.