Temperature differs from setting
ISM
sensor communication error
®
Self-test failure
Wrong settings
Out of range
Wrong password
Passwords do not match
Program memory error
40 Maintenance
Description and Resolution
ATC measured temperature differs by more than 0.5°C
from the user-defined value/temperature range.
Keep the buffer/standard temperature within the
range.
Change the temperature setting.
Data has not been transferred correctly between ISM
sen s or and meter. Reconnect the ISM
again.
Self-test has not been completed within 2 minutes or me-
ter is defective.
Restart self-test and finish within 2 minutes.
Contact METTLER TOLEDO service if problem per-
sists.
Entered value differs by less than 1 pH unit/5°C from oth-
er preset values.
Enter a higher/lower value in order to get a bigger
dif f er e nce.
Either entered value is out of range.
Enter a value which is within the range shown on
dis p lay.
or
Measured value out of range.
Make sure the electrode wetting cap has been re-
moved and that the electrode is properly connected
and placed in the sample solution.
If no electrode is connected, put the shorting clip in
the socket.
The entered PIN is not correct.
Re-enter the PIN.
Reset to factory settings, all data and settings will
be lost.
The confirmation PIN does not match with the entered
PIN.
Reenter PIN.
Meter recognizes internal error during start-up.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem per-
sists.
®