E
1.

Explanation of symbols

0123
This product complies with the Council Directive 93/42/EC from
5 September 2007 regarding medical devices, which became
effective on 21 March 2010 and bears the mark CE 0123 (TÜV
SÜD Product Service GmbH).Units with CE marks are subject to
quality inspections in accordance with this Directive and provide a
higher level of accuracy than previous calibration.
Degree of protection against electric shock: TYP BF
Please observe instructions for use
Storage and transportation conditions
Ambient temperature -5 to +50 °C
Relative humidity up to 85%,
non-condensing
Keep dry
Manufacturer
2.

Disposal

Batteries and technical appliances must not be disposed of with
domestic waste, but should be handed in at the appropriate
collection and disposal points.
EN-54

General provisions

General provisions
3.

Electromagnetic compatibility (EMC)

Technical description
The unit satisfies the EMC requirements of the international standard IEC60601-1-
2. The requirements are satisfied under the conditions described in the tables be-
low. The unit is an electrical medical product and is subject to special precautionary
measures with regard to EMC which must be published in the instructions for use.
Portable and mobile HF communications equipment can affect the unit. Use of the
unit in conjunction with non-approved accessories can affect the unit negatively
and alter the electromagnetic compatibility.The unit should not be used directly
adjacent to or between other electrical equipment.
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The device uses RF energy only for its internal
CISPR 11
function. Therefore, ist RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
Class A
emissions IEC
61000-3-2
Voltage fluctu-
Complies
ations/Flicker
emissions IEC
61000-3-3