What to do, if...
... the display remain dark:
In this case, the interface is not supplied
with power.
Please check if the power supply is con-
nected properly to power socket and
interface.
... no data logger is identified:
Register the data logger to the base
station via software Winlog.wave or
Winlog.web.
Check the base station by running the
hardware check.
Check if the antenna is connected
properly.
Check if the data loggers are still in the
reception range of the base station. If
necessary, please reduce the distance
between logger and base station or
use the Ground Plane Antenna AL 116.
... the hardware check fails:
Re-start the interface: disconnect and
then re-connect the power supply.
Run the hardware check again. If the
hardware check fails again, please
make a note of the error code and get
in contact with the ebro support.
12/2016 ba80018def04_EBI-IF-400
Français
Que faire, quand...
... l'affichage ne s'allume pas:
En ce cas-là l'interface n'est par fournie
avec courant.
Veuillez vérifier si l'alimentation élec-
trique est connectée correctement
avec la prise et l'interface.
...aucun enregistreur n'est trouvé:
Connectez les enregistreurs à l'inter-
face en utilisant le logiciel Winlog.wave
ou Winlog.web.
Vérifiez l'interface en exécutant un test
de système.
Vérifiez si l'antenne est connectée cor-
rectement.
Vérifiez si les enregistreurs se trouvent
encore dans la plage de réception de
l'interface. Si nécessaire veuillez réduire
la distance entre l'enregistreur et l'IF
400 ou utilisez l'antenne Ground-Plane.
...le test de système indique une erreur:
Redémarrez l'interface en débranchant
et embranchant l'alimentation élec-
trique.
Exécuter le test de système encore une
fois. Si le message d'erreur continue à
apparaître veuillez prendre note du
numéro d'erreur et contactez ebro.