English
What to Do, if...
...
there
is
transmission to the data logger:
- Check that the data logger's wireless
mode is active. To do this, place the
data logger close to the interface and
call the appropriate function in the
programming software.
- If necessary, use the programming
software to switch the data logger to
wireless mode.
- If the data logger does not react to
this, check if the battery is fl at and
needs to be replaced (see below).
... the data logger cannot be
contacted from the interface:
- Check that the interface has been
connected correctly and that the
evaluation software has recognized
it.
- Open the data logger (see also the
Figures
in
Battery").
- Check to see whether condensation
has penetrated the data logger
housing.
- Either wipe the condensation from
the housing or allow it to air dry by
leaving the data logger housing
open.
BA80009def04
no
wireless
the
"Replacing
Français
Que faire, quand...
... le transfert de données par
radio
vers
fonctionne pas :
- Vérifiez que le mode radio de
l'enregistreur est activé. Pour ce
faire,
proximité de l'interface et activez la
fonction correspondante du logiciel
de programmation.
- Le
cas
l'enregistreur en mode radio à l'aide
du logiciel de programmation.
- Si l'enregistreur ne réagit pas,
vérifiez si la batterie n'est pas vide et
ne doit pas être remplacée (cf. plus
bas).
... l'enregistreur ne réagit pas au
niveau de l'interface :
- Vérifiez
correctement branchée et qu'elle a
bien été reconnue par le logiciel
d'exploitation des données.
the
- Ouvrez l'enregistreur (voir aussi les
figures
"Remplacement de la batterie".
- Vérifiez s'il n'y a pas de l'humidité
dans le boîtier de l'enregistreur.
- Si oui, enlevez l'humidité contenue
dans
l'enregistreur ouvert à l'air libre
jusqu'à ce qu'il soit sec.
l'enregistreur
placez
l'enregistreur
échéant,
que
l'interface
dans
le
paragraphe
le
boîtier
ou
ne
à
basculez
est
laissez
31