Proservo NMS80
Endress+Hauser
Numéro de
Texte court
diagnostic
826
Sensor temperature
826
Sensor temperature
844
Process value out of
specification
844
Process value out of
specification
903
Current loop 1 ... 2
904
Digital output 1 ... 8
941
Echo lost
942
In safety distance
943
In blocking distance
950
Advanced diagnostics
961
Alarm 1 ... 4 HighHigh
962
Alarm 1 ... 4 High
963
Alarm 1 ... 4 Low
964
Alarm 1 ... 4 LowLow
965
Alarm 1 ... 4 HighHigh
966
Alarm 1 ... 4 High
967
Alarm 1 ... 4 Low
968
Alarm 1 ... 4 LowLow
970
Overtension
971
Undertension
1)
Le comportement de diagnostic peut être modifié.
Diagnostic et suppression des défauts
Mesures correctives
1. Check ambient temperature
2. Check process temperature
1. Check process value
2. Check application
3. Check sensor
1. Check device configuration.
2. Check wiring.
1. Check device configuration.
2. Check wiring.
1. Check process value
2. Check application
3. Check sensor
1. Check level
2. Check safety distance
3. Reset self holding
Reduced accuracy
Check level
Maintain your diagnostic event M
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check level
2. Check configuration settings
1. Check displacer and process
conditions
2. Release overtension
Check displacer and process.
Signal
Comportement du
d'état
diagnostic
[au
[au départ usine]
départ
usine]
S
Warning
F
Alarm
1)
S
Warning
S
Warning
F
Alarm
F
Alarm
S
Warning
S
Warning
S
Warning
Warning
C
Warning
C
Warning
C
Warning
C
Warning
F
Alarm
F
Alarm
F
Alarm
F
Alarm
C
Alarm
C
Alarm
147