Endress+Hauser RIA452 Mode D'emploi page 133

Masquer les pouces Voir aussi pour RIA452:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RIA452
Function
(menu
item)
Dec. point
XXXXX
XXXX.X
XXX.XX
XX.XXX
X.XXXX
Min. value
0..99999
Max. value
0..99999
Reset min
No
Yes
Reset max
No
Yes
6.3.9 Linearization table - LIN. TABLE/M21
To linearize input variables, a linearization table can be saved in the measuring instrument, e.g.
to correct the level signal of a container for volume display.
Function
(menu
item)
Counts
2..32
2
Dimension
XXXXXXXXX
XXXXX
Dec. Y value
XXXX.X
XXX.XX
XX.XXX
X.XXXX
No
Del. points
Yes
Endress+Hauser
Parameter setting
Parameter setting
Description
Number of digits after the decimal point for the
min/max value memory.
Displays the current minimum value in the memory.
Displays the current maximum value in the memory.
Resets the minimum value memory.
Resets the maximum value memory.
Description
Number of support points needed. At least two points
have to be entered.
Select the dimension from the list or dimension as free
text (max. 9 characters long).
Decimal point position for the Y-values in the
linearization table.
Delete all programmed support points.