ResMed Stellar Serie Guide De L'utilisateur page 30

Masquer les pouces Voir aussi pour Stellar Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Problem / possible cause
LCD: Low SpO
!
2
SpO
has dropped below pre-set
2
alarm level.
LCD: SpO
finger sensor failure!
2
The finger sensor is not
connected properly or delivers
faulty values.
LCD: Xpod oximeter disconnected!
The pulse oximeter is
disconnected.
LCD: Low FiO
Level!
2
FiO
has dropped below the pre-
2
set alarm level.
LCD: High FiO
Level!
2
FiO
has exceeded the pre-set
2
alarm level.
LCD: FiO
sensor disconnected!
2
The FiO
monitoring sensor is not
2
connected.
LCD: Keypad Failure!
One of the keys was held down
for more than 10 seconds or got
stuck.
LCD: Attention High Temp [42, 43, 44, 45]!
The temperature inside the
device is high.
LCD: Internal Battery Use!
The device is using the internal
battery.
LCD: External DC Power Use!
The device is powered by an
external battery.
Action
Check the attachment of the sensor.
If the problem persists, contact your clinician.
Check if the finger sensor is attached properly to the finger and
connection to the pulse oximeter.
Check if the pulse oximeter is connected properly to the device.
Perform FiO
sensor calibration.
2
If the problem persists, contact your clinician.
Perform FiO
sensor calibration.
2
If the problem persists, contact your clinician.
Connect the FiO
monitoring sensor or replace it with a new one.
2
Remove any blockages from the keypad.
Ensure the ambient temperature conditions are within the specified
operating range.
Check if the power cord is properly connected to the device if you
want to run from mains power.
Press the Alarm mute button
Check if the AC power cord is properly connected to the device if you
want to run from mains power.
Note: The alarm will be cleared automatically after one minute.
to cancel the alarm.
English
27

Publicité

Table des Matières
loading

Table des Matières