Installation / Replacement Of Batteries; Cleaning - FIM MEDICAL TABATABA V2 Manuel Utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4.1.

Installation / replacement of batteries

The TABATABA® V2 CO detector operates with two 1.5V AA batteries.
Use a Phillips screwdriver to unscrew the battery compartment cover support screw
Insert the batteries following the polarity indicators
When changing the batteries, remove the batteries and recycle them
Replace the battery compartment cover
Screw in the battery compartment cover support screw
Battery compartment
cover
4.2.

Cleaning

The device housing should be cleaned after each use with a damp cloth and a bactericide-fungicide solution.
FIM MEDICAL recommends the use of Bactinyl® odourless cloths. The use of spray is not recommended as a
misdirected jet could damage the sensor.
The trapper is single-use and should be changed systematically after each patient.
The pipette is single-use and should be changed systematically after each patient.
Warning:
Risk of cross contamination if disposable consumables are reused
Warning:
Do not sterilize the device or the accessories.
Warning:
Used or dirty accessories must be disposed of with biological waste (DASRI).
Ce document est la propriété de la Société FIM MEDICAL. Il ne peut être reproduit ou communiqué sans son autorisation.
This document is the property of FIM MEDICAL. Its contents cannot be reproduced or divulged without the company's approval
Screw
Polarity indicators
Page 41 / 43

Publicité

Table des Matières
loading

Table des Matières