4.1.
Installation / replacement of batteries
The TABATABA® V2 CO detector operates with two 1.5V AA batteries.
•
Use a Phillips screwdriver to unscrew the battery compartment cover support screw
•
Insert the batteries following the polarity indicators
When changing the batteries, remove the batteries and recycle them
•
Replace the battery compartment cover
•
Screw in the battery compartment cover support screw
Battery compartment
cover
4.2.
Cleaning
The device housing should be cleaned after each use with a damp cloth and a bactericide-fungicide solution.
FIM MEDICAL recommends the use of Bactinyl® odourless cloths. The use of spray is not recommended as a
misdirected jet could damage the sensor.
The trapper is single-use and should be changed systematically after each patient.
The pipette is single-use and should be changed systematically after each patient.
Warning:
Risk of cross contamination if disposable consumables are reused
Warning:
Do not sterilize the device or the accessories.
Warning:
Used or dirty accessories must be disposed of with biological waste (DASRI).