3.4.3. Alarm mode

Description
In this mode, a sound and visual alarm is set off when the preset threshold (during set-up of the device) has
been reached.
The TABATABA® V2 has two alarm thresholds:
• Low threshold alarm
• High threshold alarm
When an alarm threshold is reached, the device sets off an intermittent sound and visual alarm (red LED). The
alarm signal remains active for at least 10 seconds.
These two thresholds are differentiated by a more rapid frequency signal for the high level.
Use
1. Do not use CO trapper or pipette.
2. Place the TABATABA® V2 cursor to Alarm mode.
3. Press the On/Off button. The TABATABA® V2 displays "ALAR". It then goes to low consumption mode
(displayer blanks).
4. A measurement is taken every 20 seconds. A short flash is emitted by the green LED if the batteries have
sufficient power.
5. Another press on the On/Off button reactivates the device, which displays "Init" then takes an
instantaneous measurement and displays it for 20 seconds. The device then returns to low
consumption.