3.2.Settings
Press "OK/SETUP" to access TABATABA® V2 set-up in Expired mode. Parameter settings:
•
Date and time
•
Monitoring time
•
Two alarm thresholds
Use the "+" and "-" buttons to choose parameters from the scrolled messages: DATE, MONI and ALAR then press "OK/SETUP". Each of these parameters
is set using the "+" and "-" buttons and validated with "OK/SETUP".
•
Order of time and date setting: day, month, year, hours, minutes
•
Monitoring time setting offers choices of 15mins, 1h, 8h or 24h
•
Order of alarm settings: low threshold alarm, then high threshold alarm
3.3.Installation of accessories
3.3.1.Single-use trapper
Check the no-return membrane is present inside the trapper.
No-return membrane
Warning:
If the trapper is not changed and reused between two patients, there is no risk of cross-contamination (bacterial or viral
contamination) because the trap is equipped with a non-return valve. For hygienic reasons, the trappers are single-use. Systematically change
the soiled trap between each patient.
Warning:
Used or dirty accessories must be disposed of with biological waste (DASRI).
Insert the trapper into the intended position in the housing.
3.3.2.Single-use pipette
Warning:
The pipettes are single use. Systematically change the pipette between each patient. If the pipette is not changed and re-used between
two patients, a risk of cross-contamination (bacterial or viral contamination) is possible during insertion.
Warning:
Used or dirty accessories must be disposed of with biological waste (DASRI).
Ce document est la propriété de la Société FIM MEDICAL. Il ne peut être reproduit ou communiqué sans son autorisation.
This document is the property of FIM MEDICAL. Its contents cannot be reproduced or divulged without the company's approval.
V02.07.00
Août/August 2021