Use
1. Place a new single-use CO trapper into the TABATABA® V2. Check the no-return membrane is present
inside the trapper.
2. Take a new single-use pipette and check it for visual defects and foreign bodies. Place the pipette inside
the trapper and check it fits tightly into the trapper.
3. Press On/Off button.
The device displays a series of messages. Follow the instructions:
pipette in the mouth and breathe out slowly and completely.
4. Wait approximately 15 seconds, while commenting on the rising CO, to make the patient aware.
Between two measurements in Expired mode
1. Discard the used pipette.
2. Remove the used trapper and directly discard it.
3. Switch off the device.
4. Use a dry cloth to clean the opening to remove humidity.
5. Ventilate the sensor and wait one or two minutes.
6. Reinsert a new single-use trapper. Check the trapper is conform (presence of no-return membrane).
Ce document est la propriété de la Société FIM MEDICAL. Il ne peut être reproduit ou communiqué sans son autorisation.
This document is the property of FIM MEDICAL. Its contents cannot be reproduced or divulged without the company's approval
When the device displays "Insp" ask the patient to breathe in.
When the device displays "Bloc" ask the patient to hold their breath.
When the green LED lights up (device displays "expi") tell the patient to place the