7. Take a single-use pipette and check there are no defects or foreign bodies.
Insert the pipette into the trapper and check it fits tightly into the trapper. The device is ready for a new
measurement. Remind the patient not to inhale in the pipette.

3.4.2. Monitoring mode

Description
This mode records the level of atmospheric carbon monoxide over a time defined in the TABATABA® V2 set-up.
Monitoring can be run for a recording time of:
1/4 hour
1 hour
8 hours
24 hours
In all cases, a measurement is taken every 1 minute. After each recording, the TABATABA® V2 displays the
following values:
TWA (Time Weighted Average)
STEL (Short Term Exposure Limit)
Maximum value measured (C)
Use
1. Do not use CO trapper or pipette.
2. Place the TABATABA® V2 cursor on Monitoring mode (M) and switch on via the On/Off button.
3. While pressing on the Unit/0 button, press the On/Off button.
4. The "Init" message appears, and recording begins.
5. Press the On/Off button again and the TABATABA® V2 displays "RUN".
As soon as recording begins, the TABATABA® V2 performs an initial measurement, then switches off to save the
batteries. Every minute, the screen displays "RUN" and the measured value, then goes on standby. The
TABATABA® V2 can store two monitorings. If a monitoring has already been recorded, the TABATABA® V2
successively displays the corresponding TWA, STEL and C values. While pressing on the Unit/0 button, press the
On/Off button to run a new monitoring.
The TABATABA® V2 can store two monitorings, the TWA, STEL and C values displayed corresponding to the
values of the last monitoring recorded.
If the storage memory is full, the "Full" message is displayed for a few seconds. A long press on the Unit/0
button deletes all monitorings. The "CLR" message appears, indicating that the memory has been deleted.
A monitoring in progress can be stopped by activating the device (On/Off button). The screen displays "RUN".
Validate by pressing the OK/SETUP button.
Ce document est la propriété de la Société FIM MEDICAL. Il ne peut être reproduit ou communiqué sans son autorisation.
This document is the property of FIM MEDICAL. Its contents cannot be reproduced or divulged without the company's approval