ENGLISH
14
Setting osmosis system
The autoclave is also designed for loading demineralised water through an external
osmosi demineralisation system (optional accessory).
The operator before installing the system, must schedule the autoclave following the
instructions below:
Turn the autoclave on by pressing the mains switch (Fig. A-pos.11).
When the autoclave is brought up on the operating screen, press
menu SETUP OSMOSIS SYSTEM.
Activate by setting the USE OSMOSI'S SYSTEM item to ON.
Exit the menu and return to the operating screen.
Connected by a demineraliser, if the maximum water level has not been
NOTE
reached, the operation of the autoclave will be blocked.
WARNING:
The number on the osmosi management screen indicates how many STERILISATION
cycles have been performed since changing filters.
When you reach the maximum number of cycles, a message on the display will inform the
user. It's necessary to remember to reset the cycles counter when you change the filter on
the osmosis system selecting RESET OSMOSI COUNTER in the SETUP OSMOSIS
SYSTEM submenu.
Connection of the demineralizer
Turn off the autoclave, if ON (Fig.A–pos.11).
• Close the cock upstream the demineralizer;
• Install the demineralizer as indicated in its manual;
• Wrap the male thread of the hose coupler with teflon or another component ensuring
water tightness;
• Screw the hose coupler on the female thread of the clean water discharge (FIG.A–
pos.04);
• Insert the hose from the demineralizer into the hose coupler and screw it to the
autoclave;
• Insert the feeding coupler of the demineralizer into the port (FIG.A–pos.01) on the rear
panel of the autoclave;
• Open the valve upstream the demineralizer;
• Be sure there is no water leakage;
34
INSTALLATION OSMOSIS SYSTEM
Setup
and select the