7. Maintenance and calibration

To guarantee the accuracy of the multimeter over an extended
period of time, it should be calibrated every twelve months
(calibration service available at Service 2000).
Changing a fuse is described under 3. 3. (Safety requirements).
Battery changing can be found under 5.1.1.
To clean the instrument or the display window, use a clean, lint-
free, anti-static, dry cleaning cloth.
Attention!
Use no carbonaceous cleaners, petrols, alcohols or similar for
cleaning. Otherwise the surfaces of the clip adapters / test leads
can be attacked. In addition, the vapours are explosive and
damaging to health. Do not use any sharp-edged tools, screw
drivers, metal brushes or similar for cleaning.

8. Technical data and measuring tolerances

8.1 Technical data
Display .................................: 3
Maximum measuring rate ...: 2.5 measurements per second
Max. input current AC/DC ...: 20 A
40
/
-position LCD display up to
3
4
4000 (3999), with automatic pola-
rity indication
4
/
-position for frequency measure-
1
2
ments
4
/
-position "small" display
1
2
with analogue bar graph display
(bar segment display)
Input impedance ..................: 10 MΩ in parallel with 100pF (VDC
Overload indicator ...............: "OL" is displayed for overload
Operating temperature........: 0°C to +40°C, with relative humidi-
Storage temperature............: -10°C to +50°C, with relative humi-
Temperature for guaranteed
accuracy.................................: +23°C ±5K
Battery change indicator .....: "
Battery type ..........................: NEDA 1604 9V or 6F22 9V, alkaline
Weight...................................: approx. 364 g (with 9-V battery)
Dimensions (L X W X H).......: 189 x 90 x 38 mm
8.2 Measurement tolerances
Statement of accuracy in ± (% of the reading + number of digits).
Accuracy 1 year at a temperature of +23°C ± 5°C, with a relative
humidity of less than 75 %, non-condensing. The warm-up time
is 1 minute.
and VAC)
ty < 75%, non-condensing
dity < 75%, non-condensing, bat-
tery not installed
+ -
" is displayed
type
41