Models With Rs232 Interface - KERN MPC M Serie Mode D'emploi

Masquer les pouces Voir aussi pour MPC M Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12.2.2 Models with RS232 interface

Function
Settings
oFF 0*
oFF 3
Automatic cutout
oFF 5
Auto Off
oFF 15
oFF 30
oFF*
Prt
Background
illumination
Pr ACC
of display
bl on
bl oFF
Background
illumination
bl AU*
of display
1.
Interface parameter
Select desired mode by
P Prt
P Cont
Series
ASK
P cnt 2
P Stab
P Auto
2.
The currently set baud rate (b xxx) will be shown after the RS-232 mode was
confirmed. Select desired Baudrate by pressing
Available Baud rate: 600, 1200, 2400, 4800, 9600
40
Description
Automatic shutdown off
Automatic shutdown after 3 min.
Automatic shutdown after 5 min.
Automatic shutdown after 15 min.
Automatic shutdown after 30 min.
Not documented
Back lighting for display on
Display background illumination off
Backlighting for display will come on automatically as
soon as the weighing scale is operated.
RS-232 mode
Weight will be added to summation memory and printed
after pressing PRINT
Continuous data output
Not documented
Remote control instructions:
W: Send all weighing details
S: Send stable weight value
T: Taring
Z: Zeroing
Not documented
Automatic data output of stable weighing values
Weighed result will be added automatically to summation
memory and issued
Baud rate
MPC_M / MPE_HM / MPE_PM-BA-e-1414
, then confirm with
.
and confirm by
.

Publicité

Table des Matières
loading

Table des Matières