12 The menu
Access to service menu „tCH"is locked in verified weighing scales.
To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see chap. 16.
Attention:
After destruction of the seal the weighing system must be re-verified
by an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.
12.1 Navigation in the menu
Call up menu
Select function
Change settings
Exit menu/
Return to weighing
mode
38
In weighing mode, press
oFF] will be displayed.
With help of
, the individual functions can be
selected one after the other.
Confirm selected function by pressing
(MPC). The current setting will be displayed.
To select the desired setting, press
your selection by
weighing scale will return to the menu.
Press
and the scales will return to weighing mode.
MPC_M / MPE_HM / MPE_PM-BA-e-1414
and the first function [F1
(MPE) or
and confirm
(MPE) or
(MPC); the