Standard deviation can be expressed as a relative value (CV) CV = 100% x S / V
When the Finnpipette FOCUS is not in use, make sure it is stored in an upright position. We
recommend a Finnpipette stand for this purpose.
The part # refer to exploded views beginning at page 48 and 52 Standard model.
Daily checking
The pipette should be checked at the beginning of each day for dust and dirt on the outside
surfaces of the pipette. Particular attention should be paid to the tip cone. No other
solvents except 70 % ethanol should be used to clean the pipette.
Short-term service
If the pipette is used daily it should be checked and lubricated every three months.
Pipettes up to 1000 µl
1. Remove the lower tip ejector sleeve 13 by pulling it out (snap fitting).
2. Push down the ejector button and take the upper ejector top 11
between fingers.
3. Release the ejector button to up position, but hold the tip ejector top
11 in down position with the fingers.
4 Remove the tip cone by pulling it out (snap fitting).
5. Pull out the piston 14 and the spring 15 but not spring 16.
6. Clean the piston 14 and the piston spring 15 with a dry napless cloth.
7. Check the tip cone for foreign particles.
8. Grease the cleaned parts with the lubricant that comes with the pipette.
9. Reassemble the pipette components in reverse order.
5 ml and 10 ml pipettes / 2-10 ml pipettes
1. Pull down the tip ejector sleeve.
2. Note that the tip ejector button is in fully up position.
3. Remove the tip ejector by pulling it out.
4. Remove the ejector sleeve by pulling it to two parts
(snap fitting).
5. Remove the cylinder 14 by pushing the piston to the
down position and push the snap joints to release and
remove the cylinder 14.
6. Push the piston forward and clean it with a dry
napless cloth.
7. Check the tip cone for foreign particles.
8. Grease the cleaned parts with the lubricant that comes
with the pipette.
9. Reassemble the pipette components in reverse order.
Long-term service
If the pipette is used daily it should be serviced every six months.
Pipettes up to 1000µl
1. Remove the lower tip ejector sleeve 13 by pulling it out (snap fitting).
2. Push down the ejector button and take the upper ejector top 11 between fingers.
3. Release the ejector button to up position, but hold the tip ejector top 11 in down
position with the fingers.
4. Remove the tip cone by pulling it out (snap fitting).
5. Pull out the piston 14 and the spring 15.
6. Remove spring hatch 25 by pressing it down and in the same time turning 90 degrees.
Take out the spring 16.
s = standards deviation
V
)
2
v = mean volume
n = number of measurements

Maintenance

9