  • FRANÇAIS, page 25
8.1
Electrical Safety
Guidance and Manufacturer's Declaration - Electromagnetic Emission
DUO is intended for use in the electromagnetic environment specified below. The user of DUO should assure that it is used
in such an environment.
Applicable Emissions Test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic Emissions IEC
Not Applicable
6100-3-2
Voltage fluctuations/flicker
Not Applicable
emissions
IEC 61000-3-3
Electromagnetic Environment- Guidance
DUO uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
DUO is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Warning: The use of accessories other than those
specified, with the exception of accessories sold by
Eko as replacement parts, may result in increased
emissions or decreased immunity of DUO.
Warning: DUO should not be used adjacent to or
stacked with other equipment. If adjacent or stacked
use is necessary, DUO should be observed to verify
normal operation in the configuration in which it will be
used.