Connecting The Connecting Cable To The Platform - Seca 674 Mode D'emploi

Masquer les pouces Voir aussi pour 674:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 81
Connecting the
connecting cable to the
60 •
platform
1. Place the scales on a firm, level surface.
2. Connect the connecting cable to the electronics
box of the scales.
NOTE
When selecting the setup location, take
account of the length of the connecting cable
so that you can place the display housing on
the wall bracket.
ATTENTION!
Incorrect measurement due to force shunts
If the cable of the display housing contacts the
weighing platform, the patient's weight cannot
be measured correctly.
– Route the cable in such a way that it cannot
contact the weighing platform.

Publicité

Table des Matières
loading

Table des Matières