Error/display
ERR1
ERR2
ERR3
ERR4
ERR5
ERR6
"Running" seg-
ments appear on
the display
No display. Indi-
cator lamps do
not light up
Cause
Heating element
has overheated,
dry measurement
Measured
temperature >
300 °C. Boil-
ing point of
the measured
sample is above
300 °C
Internal tem-
perature of the
device > 50 °C.
Measurements
not possible
Sensor head
is missing or
defective
Safety controller
has switched off
Temperature sen-
sor defective
Temperature
sensor has a
temperature >
55 °C
Connection to
the mains has
been interrupted
or fuse on the
device has blown
Remedy
Immerse the sensor head correctly in sample
fluid
Switch the device off and on again
Press the "Start" button
Only measure glycol-based brake fluids
Clean the sensor head by carrying out at least
three measuring processes with clean brake
fluid
Use a new sample beaker for each measure-
ment
Switch the device off and leave it to cool
Do not place the device on heaters or in
sunlight
If the error occurs repeatedly, send in the
device for inspection
Install or replace the sensor head
Switch the device off and then switch it on
again after 5 seconds
If the error occurs repeatedly, send in the
device for inspection
Send in the device for repair
Leave the temperature sensor to cool. When
the temperature sensors cools down to a
temperature < 55 °C, the measurement starts
automatically
Check connection to the mains
Change the fuse, see page 29