Feedwater quality
Millipore recommends using water treated by ELIX (Electrodeionisation, E.D.I.) or Reverse Osmosis (RiOs).
Environmental Conditions
Ambient storage temperature
Ambient operating temperature
Humidity
System performance
Purified water quality
Resistivity
Pyrogens
TOC*
Micro-organisms
Particles (0.22 µm)
Flowrates
Product water flowrate
Noise level in dB A at 1 metre
* Test conditions: Milli-Q system was equipped with a Q-Gard purification pack and a Quantum EX ultrapure
cartridge. The feedwater to the Milli-Q came from a RiOs Reverse Osmosis system. TOC levels in the
feedwater were < 50 ppb. The quality of the Milli-Q product water can vary as a function of the quality of
the feedwater.
Dimensions and operating weight
Height
Length
Depth
Weight(s)
INTRODUCTION
5 °C < T < 40 °C
5 °C < T < 25 °C
20 % - 80 % without condensation
18.2 MΩ
cm at 25 °C
•
0.02 EU/ml
5 - 10 ppb
1 - 5
ppb
2 - 5
ppb
≤ 1 cfu/ml
< 1/ml
Up to1.5 litre/minute
Up to1.0 litre/minute
42 dB A
(with Q-Gard and Quantum)
455 mm
255 mm
355 mm (includes the wall mounting points on rear)
16.0 kg
16.8 kg
16.3 kg
17.1 kg
+ 0.6 kg with A10 TOC module
Milli-Q
Ac Gr Bio Syn A10
9