Interrupting the sanitisation cycle of a UF cartridge
If a sanitant chemical has been introduced into the Milli-Q, it is absolutely necessary to let the sanitisation
cycle finish in order to completely rinse the sanitant from the system.
If no sanitant has been introduced, then the sanitisation cycle can be cancelled by following these instruc-
tions.
OPERATE MEASURE
CLEANING
MENU
STAND/BY
Purging trapped air from the UF cartridge
OPERATE MEASURE CLEANING
MENU
STAND/BY
OPERATE MEASURE CLEANING
MENU
STAND/BY
Interrupting the A10 cleaning cycle
If an A10 cleaning cycle has been started from the MENU service mode, then it can be cancelled at any
time by pressing on the MENU button.
Press CLEANING for 10 seconds
↓
At the end of the sanitisation cycle, the
Milli-Q will go back to its initial mode of
operation.
In either PRE OPERATE or PRODUCT
Press MENU for 2 seconds
↓
Press MENU 4 times
Open the point of use valve to begin the
AIR PURGE cycle.
Dispense product water from the system to
the drain.
At the end of the air purge, the system will
go back into STANDBY mode.
Close the point of use valve to place the
system in PRE OPERATE mode.