MAINTENANCE
Milli-Q
Ac Gr Bio Syn A10
Cleaning the screen filter in the feedwater line (Figure 3)
1. Close the valve on the feedwater line.
2. Remove the feedwater tubing (C) from the fitting (F).
3. Unscrew the fitting (E) from the feedwater pipe (D) and the other fitting (F).
4. Clean the screen filter (G).
5. Proceed in reverse order to reinstall the screen filter.
Cleaning the A10
Periodically, the detection cell in the A10 needs to be cleaned of any built-up residual organic matter. If this
build-up happens, then the displayed TOC values could be erratic or higher then previously seen.
To fix this, an autocleaning cycle is needed to oxidise any organic contaminants present in the A10 detec-
tion cell.
To perform a cleaning cycle of the A10, follow these instructions:
OPERATE
MEASURE CLEANING
STAND/BY
OPERATE
MEASURE CLEANING
STAND/BY
OPERATE
MEASURE CLEANING
STAND/BY
Note: It is possible to get water from the point of use valve during the A10 CLEANING mode, but the
flowrate is lower.
24
In PRE OPERATE mode
Press MENU for 2 seconds
MENU
Press MENU 3 times
MENU
Wait 5 seconds to start the A10 cycle.
After 60 mins. the system will auto-
matically return to its initial operating
mode.
(To interrupt the cleaning cycle)
Press MENU for 2 seconds to return to
MENU
the initial operating mode.
P R O D . T I M E S E T U P
C O U N T E R :
0 m n
S E R V I C E : A 1 0
A 1 0 C L E A N I N G
6 0
M Ω cm
1 8 . 2
T O C :
3 p p b