List of ALARM codes

When the ALARM LED is flashing, an error code number will be displayed which indicates the nature of the
problem inside the Milli-Q.
System
DESCRIPTION
Error #
1
The motor voltage is above its
recommended operating value.
2
Temperature < minimum.
3
Temperature > maximum.
4
Resistivity < minimum (off-scale).
5
Resistivity > maximum (measurement
is not representative).
6
Motor voltage error.
7
UV lamp voltage error.
8
Defective UV lamp.
9
Error with the electronics reference
signal.
10
Error in EEPROM storage.
11
Communication error with the A10
TROUBLESHOOTING
REMEDY
Contact Millipore Technical Service.
The measured temperature needs a short period of
time to stabilise. If the message persists, Contact
Millipore Technical Service .
The measured temperature needs a short period of
time to stabilise. If the message persists, Contact
Millipore Technical Service .
Resistivity of product water is off-scale.
Let the Milli-Q operate for a few minutes to force any
air out of the resistivity cell. If the message persists,
Contact Millipore Technical Service .
Resistivity of water is off-scale. Let the Milli-Q
operate for a few minutes to force any air out of the
resistivity cell. If the message persists, Contact
Millipore Technical Service .
Contact Millipore Technical Service.
Contact Millipore Technical Service.
Contact Millipore Technical Service.
Contact Millipore Technical Service.
Contact Millipore Technical Service.
Press OPERATE/STANDBY to reinitialise the Milli-Q.
If the problem persists, then contact Millipore
Technical Service.