Remove the tubing from the point of use. Remove the adapter fitting. Reinstall the MILLIPAK 40.
The Milli-Q system is now ready for normal use.
ACTION
Put the Milli-Q into STANDBY mode by
pressing OPERATE / STANDBY for
2 seconds
Remove the MILLIPAK 40 from the point of use (figures 6 and 7)
Screw the adapter fitting (figure 2, G) onto the point of use.
Fit the 12 mm OD tubing (figure 2, F) onto the end of the barb fitting. Place
the other end of this tubing into a drain or sink.
Remove the sanitisation port plug from the system. Keep it near the system.
(figure 1, L)
Introduce 3 grams of sodium hydroxide (NaOH) into the sanitisation port.
Screw the sanitisation port plug back onto the system. Check that it is secure
to avoid leaks.
Press CLEANING for 2 seconds
(Pressing CLEANING again allows the
choice of the longer sanitisation cycle).
Wait 10 seconds to validate the selec-
tion.
Bring the point of use trigger forward
to begin the sanitisation cycle.
When 400 minutes are shown, put the
trigger back to its upright position.
Note: If the valve is not closed, then the Sanitisation time will not count
down any further.
At the end of the sanitisation cycle,
the system will automatically go into
PRE-OPERA TE mode.
MAINTENANCE
DISPLAY
S T A N D B Y
C L E A N I N G :
1
C L E A N I N G :
1
O P E N T H E
V A L V E
C L E A N I N G :
4 2 1 m n
C L E A N I N G :
4 0 0 m n
C L O S E T H E
V A L V E
P R E O P E R A T E
Milli-Q
Ac Gr Bio Syn A10
23