Remove the tubing from the point of use. Remove the adapter fitting. Reinstall the MILLIPAK 40.
The Milli-Q system is now ready for normal use.
ACTION
Put the Milli-Q into STANDBY mode by
pressing OPERATE / STANDBY for
2 seconds
↓
Remove the MILLIPAK 40 from the point of use (figures 6 and 7)
↓
Screw the adapter fitting (figure 2, G) onto the point of use.
↓
Fit the 12 mm OD tubing (figure 2, F) onto the end of the barb fitting. Place
the other end of this tubing into a drain or sink.
↓
Remove the sanitisation port plug from the system. Keep it near the system.
(figure 1, L)
↓
Introduce 3 grams of sodium hydroxide (NaOH) into the sanitisation port.
↓
Screw the sanitisation port plug back onto the system. Check that it is secure
to avoid leaks.
↓
Press CLEANING for 2 seconds
(Pressing CLEANING again allows the
choice of the longer sanitisation cycle).
↓
Wait 10 seconds to validate the selec-
tion.
↓
Bring the point of use trigger forward
to begin the sanitisation cycle.
↓
When 400 minutes are shown, put the
trigger back to its upright position.
↓
Note: If the valve is not closed, then the Sanitisation time will not count
down any further.
At the end of the sanitisation cycle,
the system will automatically go into
PRE-OPERA TE mode.
↓
MAINTENANCE
DISPLAY
S T A N D B Y
C L E A N I N G :
1
C L E A N I N G :
1
O P E N T H E
V A L V E
C L E A N I N G :
4 2 1 m n
C L E A N I N G :
4 0 0 m n
C L O S E T H E
V A L V E
P R E O P E R A T E
Milli-Q
Ac Gr Bio Syn A10