APPENDIX 1
Milli-Q
Ac Gr Bio Syn A10
Replacing the main electrical power fuse
Regulating the mobility of the point of use arm
The point of use arm can be adjusted in 2 locations. This adjustment is done by tightening the arm screw(s)
with the 6 sided key.
Recycling the reject water from the A10 TOC
The reject water from the use of the A10 can be recycled to a feed reservoir. Contact Millipore Technical
Service for more details.
System is not operated for a long time
Keep the Milli-Q in PRE OPERATE mode when water is not needed. In this mode, the Milli-Q will operate all
cycles to keep the water quality optimum.
If the Milli-Q is to be shut down for a long time, contact Millipore Technical Service for further information .
30
1.
Put the system into STANDBY mode by pressing OPERATE/STANDBY for 2 seconds.
2.
Turn off the power switch (figure 1, H) by putting it into the 0 position.
3.
Unplug the electrical power cord from the wall and from the system.
4.
Remove the fuse holder (figure 1, J).
5.
Remove the blown fuse and replace it with the spare fuse.
Note: Put a new spare fuse into the holder for future use.
6.
Replace the fuse holder and plug the power cord back in to the system and to the wall.
Do not overtighten the arm screws. This may block movement of the arm or can damage the
arm itself.