Section 3.0 Description Of Components; Accessory Components; Classifications; Technical Specifications - Gyrus Acmi Berkeley VC-10 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

80807 REV KA
Berkeley VC-10 Vacuum Curettage System • Instructions for Use and Technical Manual
Section 3.0
Description of Components
3.1

Accessory Components

Quantity Description
6
Berkeley SafeTouch
Collection Canisters, single-use disposable (for 3/8" tubing sets)
®
5
Berkeley SafeTouch Tissue Traps
10
Vacurette Cannulas, 8 mm, curved, sterile, single-use, disposable
5
18" Series Connecting Hose
5
3/8" Disposable Collection Tubing Set, sterile, single-use, disposable
1
Disposa-Filter Assembly
1
Berkeley VC-10 In-Service Video
1
Instructions for Use manual
See Section 11.0 for complete list of Berkeley Vacuum Curettage accessory components.
3.2

Classifications

Type Construction:
Protection Against Electric Shock Hazards–
Class of Protection:
Degree of Protection:
Rated Operation:
3.3

Technical Specifications

(for reference only—typical measurements/performance)
Operating Voltage–
VC-10:
Power Consumption:
Fuse Ratings & Type–
VC-10:
Earth/Chassis Leakage Current:
Ground Resist:
Horsepower Rating:
Dimensions:
Weight:
Vacuum Indicator Accuracy:
Environmental Specifications–
Operating Temperature:
Storage Temperature:
Relative Humidity:
Volume Reference:
Mobile Equipment
Class-I
Type-B
Continuous
115 VAC, 60 Hz
400VA
Qty:(2) (T6.3AH 250V)
Gyrus ACMI
: P2700563-001
REF
Less than 150 μa
Less than 0.1 ohms case-to-ground connection
1/3 hp
18.3" W x 25" D x 38" H
(46.4 cm W x 65.3 cm D x 96.5 cm H)
110 lbs.
Within ± 5% of full-scale value
5–40 °C, 70 kPa to 106 kPa atmospheric pressure
0–50 °C, 60 kPa to 106 kPa atmospheric pressure
10–90% non-condensing
1 cc (cubic centimeter) = 1 ml (milliliter)