Cleaning / Disinfection
Clean on a regular basis. Clean surfaces with a damp cloth using
an appropriate pH-neutral detergent. Do not use solvents or
strong liquids, this may damage surfaces on the lifter. For disinfec-
tion when needed; use isopropyl alcohol. Avoid abrasive cleaning
products. Check emergency stop and emergency lowering after
cleaning. The lift should not be exposed to running water.
Make sure not to damage or remove labels
when cleaning.
The hoist and hand control have IP24 protec-
tion class which indicates: >12.5mm protection
against access to hazardous parts, and that
water splashed against the enclosure from any
direction shall have no harmful effects.
Caregiver and client is exposed to the follow-
ing materials: Polypropilene and polyester
(slings*), PC/ABS (hand control and hoist
cover), powder coated aluminium, powder
coated zink (sling bars).
* = Applied part
In rare cases client or caregiver exposed to
polypropilene and polyester may suffer from
allergic reactions such as skin irritation, mild
itching and skin tenderness.
Reconditioning
Follow cleaning and installation procedure, complete periodic
inspection and use checklist after installation to recondition the
lifter.
Recycling
Refer to "Recycling instructions" for how to properly dispose of
product. This can be found on www.etac.com.
Spare parts
A list of spare parts is available on request.
Periodic inspection scope
Periodic inspection is a visual examination (particularly of the
lifter's load bearing structure and lifting mechanism with
attachments, brakes, controls, safety devices and person-support
devices) according to Periodic Inspection Report for Molift Air
200 and 350.
See www.etac.com.
If there is any problems with the lifter that
could jeopardize someones safety, the lifter
shall immediately be taken out of service
and marked "out of order". Do not use the
lifter untill it is repaired. Contact you local
Molift/Etac dealer/distributer to report any
problems.
Periodic Inspection shall be performed at least once a year or
more frequently if required by local requirements. The inspection
must be performed by service personnel authorized by Etac.
Contact Etac at molift@etac.com for training and
authorization or recommendation of an approved service partner.
When performing a periodic inspection, the inspector shall fill
out the inspection report for Molift Air. The reports should be
retained by the person(s) responsible for servicing the lifter. If the
inspection reveals defects and damages, the lifter shall be taken
out of operation and marked "out of order" until it is repaired.
Troubleshooting
Symptom
The lifter does not respond
to Hand control action
The lifter does not respond
to Hand control up button
When using Electrical
Emergency lowering the
lifting band goes upwards
10
Molift Air / www.etac.com
Possible Cause/Action
Emergency Stop is activated.
Deactivate by pushing button
back in.
The lifters electronics is
overheated.
Wait for it to cool down.
Hand control is not plugged
in properly. Hand control or
plug or cord can be broken an
should be replaced.
Lifting strap angle sensor is
activated. Adjust lifting strap
or move hoist to reduce lifting
strap angle.
Lifting band has gone
completely out, and has been
winded the wrong way.
Use manual emergency lower-
ing. Send lifter to service.